RENATRON100

Description

RENATRON Reprocessing Systems, introduced in 1982, have reprocessed tens of millions of dialyzers over 40 years.

 

RENATRON SERIES SPECIFICATIONS

ELECTRICAL
RENATRON

Station

Interface ModuleLabel Printer
Voltage100, 220, 230,

240 VAC,

50/60 Hz

Single Phase

115/230 VAC

50/60 Hz

Single Phase

100-240

VAC

50/60 Hz

Consumption460 VA Max20 VA Max39.1 Watt Max
WATER REQUIREMENTS FOR A SINGLE RENATRON STATION
Feed Line Flow0.46 gallons/min. (1.75 L/Min) at average pressure of 20-55 psig
Peak Flow1.6 G/Min (6 L/Min) This peak flow is for 15-35 seconds in duration
QualityMust meet ANSI/AAMI RD-62 and RD-47
Temperature59-75°F (15-24°C)
Static PressureShould not be in excess of 55 psig (379 K Pa, 3.79 Bars)
DrainMinimum flow requirements is 1.6 G/Min (6 L/Min) per reprocessing station
ENVIRONMENTAL LIMITS
Room Temperature41-104°F (5-40°C)
Room HumidityMaximum relative humidity of 80% at 87.8°F (31°C)
SIZE AND WEIGHT
Width (Approx)Depth (Approx)Height (Approx)Weight (Approx)
RENATRON II, PA Series14.5” (41 cm)12.3” (31 cm)16” (33 cm)50 lbs (23 kg)
Interface Module12” (30.5 cm)11” (28 cm)3.5” (8.9 cm)6 lbs (2.7 kg)
Drip Tray (Single)*15.25” (39 cm)28” (71 cm)3.5” (9 cm)3 lbs (1.36 kg)
Drip Tray (Dual)*33.25” (85 cm)28” (71 cm)3.5” (9 cm)6 lbs (2.73 kg)
Label Printer7.9” (20 cm)9.75” (24.8 cm)6.8” (17.3 cm)3.2 lbs (1.5 kg)
Featuring a sturdy bracket for cord and hose management, as well as easily accessible sample ports with drip tray.Constructed of Aluminum, the ROVER ALX system is rust proof and easy to clean.